MODEL 3081 pH/ORP | SECTION 3.0 |
| WIRING |
SECTION 3.0
WIRING
3.1General Information
3.2Wiring Diagrams
3.1 GENERAL INFORMATION
pH and ORP sensors manufactured by Rosemount Analytical can be wired to the Model 3081 pH/ORP transmitter in three ways:
1.directly to the transmitter,
2.to a
3.to a remote junction box and then from the remote junction box to the transmitter. The pH (or ORP) signal can also be preamplified in one of four places.
1.in the sensor,
2.in a junction box mounted on the sensor,
3.in a remote junction box.
4.at the transmitter.
Figure 3-1 illustrates the various arrangements.
FIGURE 3-1. Wiring and Preamplifier Configurations for pH and ORP Sensors.
The asterisk identifies the location of the preamplifier. In (a) and (b) the sensor is wired directly to the transmitter. The signal is amplified at the sensor (a) or at the transmitter (b). In (c) the sensor is wired through a
14