Emerson Process Management 3081 pH/ORP instruction manual Rfh, Rwjh, Rwjl, Rfl

Models: 3081 pH/ORP

1 150
Download 150 pages 18.28 Kb
Page 77
Image 77

MODEL 3081 pH/ORP

PROGRAM

rFH

 

1

400

EXIT

NEXT ENTER

PROGRAM

rWJH

 

0

040

EXIT

NEXT ENTER

PROGRAM

rWJL

 

0

000

EXIT

NEXT ENTER

PROGRAM

rFL

 

0

000

EXIT

NEXT ENTER

SECTION 8.0

PROGRAMMING FOR pH MEASUREMENTS

12. The rFH prompt appears. Use the editing keys to change the display to the desired ref- erence electrode high impedance fault value. The allowed ranges are

Type of reference electrode

Allowed range

Low impedance (LO in step 11)

0

- 2000 kilohms

High impedance (HI in step 11)

0

- 2000 megohms

Entering 0000 disables the feature. When the reference electrode impedance goes above the fault value, the transmitter displays the diagnostic message rEFFAIL and sets a fault condition. Press ENTER to save.

13. The rWH prompt appears. Use the editing keys to change the display to the desired ref- erence electrode high impedance warning value. The allowed ranges are

Type of reference electrode

Allowed range

Low impedance (LO in step 11)

0

- 2000 kilohms

 

 

 

High impedance (HI in step 11)

0

- 2000 megohms

Entering 0000 disables the feature. When the reference electrode impedance goes above the fault value, the transmitter displays the diagnostic message rEFWArn. Press ENTER to save.

14. The rWL prompt appears. Use the editing keys to change the display to the desired ref- erence electrode low impedance warning value. The allowed ranges are

Type of reference electrode

Allowed range

Low impedance (LO in step 11)

not applicable

High impedance (HI in step 11)

0 - 900 megohms

Entering 0000 disables the feature. When the reference electrode impedance goes below the warning value, the transmitter displays the diagnostic message rEFWArn. Press ENTER to save. The prompt appears but is disabled when LO is selected in step 11.

15.The rFL prompt appears. Use the editing keys to change the display to the desired ref- erence electrode low impedance fault value. The allowed ranges are

Type of reference electrode

Allowed range

Low impedance (LO in step 11)

not applicable

 

 

High impedance (HI in step 11)

0 - 900 megohms

Entering 0000 disables the feature. When the reference electrode impedance goes below the fault value, the transmitter displays the diagnostic message rEFFAIL and sets a fault condition. Press ENTER to save. The prompt appears but is disabled when LO is selected in step 11.

16. Press EXIT to return to the process display.

63

Page 77
Image 77
Emerson Process Management 3081 pH/ORP instruction manual Rfh, Rwjh, Rwjl, Rfl