23
MODEL3081 pH/ORP SECTION 3.0
WIRING
NOTE: MODEL 328A-08 HAS UNDRESSED COAXIALCABLE. SEE SENSOR INSTRUCTION
SHEETFOR CABLE PREPARATION PROCEDURE.
FIGURE 3-17. Wire functions for Model 328A-07.
NOTES:
1. PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANS-
MITTER" POSITION (SEE SECTION 2.2).
2. JUMPERS SUPPLIED BY CUSTOMER.
3. MODEL 328A-08 HAS UNDRESSED COAXIALCABLE. CABLE
MUSTBE TERMINATED BEFORE WIRING SENSOR TO TRANS-
MITTER. CENTRALCONDUCTOR IS pH SIGNAL (TB-10) AND
SHIELD IS REFERENCE SIGNAL(TB-7). SEE SENSOR
INSTRUCTION SHEETFOR DETAILS.
4. AUTOMATIC TEMPERATURE COMPENSATION MUSTBE
TURNED OFF. SEE SECTION 8.5.
FIGURE 3-18. Wiring diagram for Model 328A.