MODEL 3081 pH/ORP

SECTION 3.0

 

WIRING

NOTE: MODEL 328A-08 HAS UNDRESSED COAXIAL CABLE. SEE SENSOR INSTRUCTION

SHEET FOR CABLE PREPARATION PROCEDURE.

FIGURE 3-17. Wire functions for Model 328A-07.

NOTES:

1.PLACE PREAMPLIFIER SELECTION SWITCH S1 IN "TRANS- MITTER" POSITION (SEE SECTION 2.2).

2.JUMPERS SUPPLIED BY CUSTOMER.

3.MODEL 328A-08 HAS UNDRESSED COAXIAL CABLE. CABLE MUST BE TERMINATED BEFORE WIRING SENSOR TO TRANS- MITTER. CENTRAL CONDUCTOR IS pH SIGNAL (TB-10) AND SHIELD IS REFERENCE SIGNAL (TB-7). SEE SENSOR INSTRUCTION SHEET FOR DETAILS.

4.AUTOMATIC TEMPERATURE COMPENSATION MUST BE TURNED OFF. SEE SECTION 8.5.

FIGURE 3-18. Wiring diagram for Model 328A.

23

Page 37
Image 37
Emerson Process Management 3081 pH/ORP instruction manual Wire functions for Model 328A-07