MODEL 3081 pH/ORP | SECTION 1.0 |
| DESCRIPTION AND SPECIFICATIONS |
1.4 SPECIFICATIONS - pH | 1.5 SPECIFICATIONS - ORP |
pH Input Range: 0 to 14 pH
Temperature Input Range: 5°F to 248°F
Output Scale Expansion: Continuously expandable between pH 0 and 14
Accuracy at 25°C: ±0.01 pH
Repeatability at 25°C: ±0.01 pH
Resolution: 0.01 pH and 0.1°C or °F
Stability at 25°C: 0.25% per year
ORP Input Range:
Temperature Input Range: 5°F to 248°F
Output Scale Expansion: Continuously expandable between
Accuracy at 25°C: ±1 mV
Repeatability at 25°C: ±1 mV
Resolution: 1 mV and 0.1°C or °F
Stability at 25°C: 0.25% per year
Temperature Compensation: Automatic or manual between 5°F to 248°F
Solution Temperature Compensation: Transmitter will convert pH measured at any temperature to the pH at 25°C. Temperature coefficient is programmable between
Calibration: Automatic
JIS 8802, BSM, Merck, and Ingold buffers.
1.6 ORDERING INFORMATION
The Model 3081 pH/ORP Smart
MODEL
3081pH/ORP HART SMART TWO-WIRE MICROPROCESSOR TRANSMITTER
Code | REQUIRED SELECTION |
|
LCD (Infrared Remote Control - included) | ||
|
|
|
LCD (Infrared Remote Control - not included) | ||
|
| |
Code | AGENCY APPROVALS |
67FM approved, intrinsically safe when used with approved sensor and safety barrier,
69CSA approved, intrinsically safe when used with approved sensor and safety barrier,
73CENELEC approved, intrinsically safe, safety barrier required
3081pH - 01 - 20 - 67 | EXAMPLE |
4