MODEL 3081 pH/ORP | SECTION 10.0 |
| PROGRAMMING FOR ORP MEASUREMENTS |
10.6 DISPLAY UNITS
10.6.1 Purpose
This section describes how to do the following:
1.switch the process display units between pH and ORP (millivolts),
2.select °C or °F for the temperature display,
3.select percent of full scale or milliamps for the output display,
4.program a security code.
10.6.2 Definitions
1.DISPLAY UNITS. Select pH if the transmitter is being used to measure pH. Select ORP if the transmitter is being used to measure ORP. The units selected are shown in the main display next to the measured value.
2.OUTPUT CURRENT DISPLAY. The transmitter generates a 4 to 20 mA output signal directly proportional to the ORP of the sample. The output signal also appears on the
3.SECURITY CODE. The security code unlocks the transmitter and allows complete access to all menus. The transmitter is shipped with security code disabled.
PROGRAM
dISPLAY
EXIT NEXT ENTER
PROGRAM
tYPE |
| PH |
| ||
EXIT |
| ENTER | |||
|
|
|
|
| |
|
|
|
|
| |
PROGRAM |
|
|
|
| |
tEMP |
|
| |||
C | |||||
EXIT |
| ENTER | |||
|
|
|
|
|
|
|
|
|
|
|
|
PROGRAM |
|
|
|
| |
OUtPUt |
|
|
|
|
|
| CUR |
| |||
EXIT |
|
| ENTER |
10.6.3 Procedure
1. Press PROG on the infrared remote controller (IRC).
2. Press NEXT until the dISPLAY
3. The screen displays the tYPE prompt. Press Ï or Ð to toggle between pH and OrP. Press ENTER to save.
4. The screen displays the tEMP prompt.Press Ï or Ð to toggle between C and F. Press ENTER to save.
5. The screen displays the OUtPUt prompt. Press Ï or Ð to toggle between % and CUr. Press ENTER to save.
PROGRAM
CODE
EXIT
000
ENTER
6. The screen displays the COdE prompt. Use the editing keys to enter a security code between 001 and 999. Entering 000 disables the security feature. Press ENTER to save.
7. Press EXIT to return to the process display.
83