MODEL 3081 pH/ORP | SECTION 10.0 |
| PROGRAMMING FOR ORP MEASUREMENTS |
SECTION 10.0
PROGRAMMING FOR ORP MEASUREMENTS
10.1General
10.2Entering and Leaving the Program Menu
10.3Output Ranging
10.4Diagnostic Parameters
10.5Temperature Element
10.6Display Units
10.7Generating a Test Current
10.1 GENERAL
This section describes how to do the following:
1.assign ORP values to the 4 and 20 mA outputs,
2.set the current generated by the transmitter during hold,
3.set the current generated by the transmitter when a fault is detected,
4.change sensor diagnostic limits,
5.change the units of the displayed variables,
6.program a security code,
7.simulate output currents for testing.
Factory default settings are given in Table
10.2 ENTERING AND LEAVING THE PROGRAM MENU
Press PROG on the infrared remote controller (IRC) to enter the Program menu. To save new settings, press ENTER . To leave the Program menu without saving new values, press EXIT . Pressing EXIT with a prompt showing returns the display to the first prompt in the
If program settings are protected with a security code, pressing PROG or CAL will cause the Id screen to appear. Key in the security code and press ENTER . The first
A transmitter adjacent to the one being programmed may pick up signals from the IRC. To avoid accidentally changing settings, use a different security code for each nearby transmitter. See Section 5.7, Security, and Section 10.6, Display Units, for details.
75