Empty Pipe

Set the test criteria for the Empty Pipe condition. The factory default for this value is set to three percent with limits configurable between one and ten percent.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, Set Pass/Fail Criteria, Empty Pipe Limit

8714i

8714i Meter Verification Test Scope

The 8714i Meter Verification can be used to verify the entire flowmeter installation, or individual parts such as the transmitter or sensor. This parameter is set at the time that the 8714i Meter Verification test is initiated.

All

Run the 8714i Meter Verification test and verify the entire flowmeter installation. This parameter results in the verification test performing the transmitter calibration verification, sensor calibration verification, coil health check, and electrode health check. Transmitter calibration and sensor calibration are verified to the percentage associated with the test condition selected when the test was initiated.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, Run 8714i

Context Menu, Diagnostics and Tests, 8714i Meter Verification

Transmitter

Run the 8714i Meter Verification test on the transmitter only. This results in the verification test only checking the transmitter calibration to the limits of the test criteria selected when the 8714i Meter Verification test was initiated.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, Run 8714i

Context Menu, Diagnostics and Tests, 8714i Meter Verification

Sensor

8714i Meter Verification Test Results Parameters

Run the 8714i Meter Verification test on the sensor only. This results in the verification test checking the sensor calibration to the limits of the test criteria selected when the 8714i Meter Verification test was initiated, verifying the coil circuit health, and the electrode circuit health.

375Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, Run 8714i

AMS Tab Context Menu, Diagnostics and Tests, 8714i Meter Verification

Once the 8714i Meter Verification test is initiated, the transmitter will make several measurements to verify the transmitter calibration, sensor calibration, coil circuit health, and electrode circuit health. The results of these tests can be reviewed and recorded on the calibration verification report found on page C-16.This report can be used to validate that the meter is within the required calibration limits to comply with governmental regulatory agencies such as the Environmental Protection Agency or Food and Drug Administration.