Reference Manual

00809-0100-4663, Rev BA January 2010

FOUNDATION™ FIELDBUS SPECIFICATIONS

Table A-1. Block Information

Schedule Entries

Seven (7)

Links

Twenty (20)

Virtual Communications Relationships (VCRs)

One (1) predefined (F6, F7) Nineteen (19) configurable (see Table 1)

Block

Execution Time (Milliseconds)

Resource (RB)

—

Transducer (TB)

—

Analog Input (AI)

10

Proportional/Integral/

10

Derivative (PID)

 

Integrator (INT)

10

Arithmetic (AR)

10

Reverse Flow

Detects and reports reverse flow

Software Lockout

A write-lock switch and software lockout are provided in the resource function block.

Turn-on Time

5 minutes to rated accuracy from power up; 10 seconds from power interruption.

Start-up Time

50 ms from zero flow.

Low Flow Cutoff

Adjustable between 0.01 and 38.37 ft/s (0.003 and 11.7 m/s). Below selected value, output is driven to the zero flow rate signal level.

Overrange Capability

Signal output will remain linear until 110% of upper range value or 44 ft/s (13 m/s). The signal output will remain constant above these values. Out of range message displayed on local display and field communicator.

Damping

Adjustable between 0 and 256 seconds.

Sensor Compensation

Rosemount sensors are flow-calibrated and assigned a calibration factor at the factory. The calibration factor is entered into the transmitter, enabling interchangeability of sensors without calculations or a compromise in standard accuracy.