Rosemount 8732

Reference Manual

00809-0100-4663, Rev BA January 2010

Table 5-3. Flange Bolt Torque and Bolt Load Specifications for Rosemount 8705

 

 

 

 

 

 

Polyurethane Liner

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Size

 

 

PN 10

PN 16

 

 

PN 25

PN 40

 

 

 

 

 

 

 

 

 

 

 

 

 

Code

Line Size

(Newton-meter)

(Newton)

(Newton-meter)

(Newton)

(Newton-meter)

(Newton)

(Newton-meter)

 

(Newton)

005

1/2-inch (15 mm)

1

 

521

1

826

2

 

1293

6

 

3333

010

1 inch (25 mm)

2

 

1191

3

1890

5

 

2958

10

 

5555

015

11/2 inch (40 mm)

5

 

1960

7

3109

12

 

4867

20

 

8332

020

2 inch (50 mm)

6

 

2535

10

4021

15

 

6294

26

 

10831

030

3 inch (80 mm)

5

 

2246

9

3563

13

 

5577

24

 

19998

040

4 inch (100 mm)

7

 

3033

12

4812

23

 

7531

35

 

11665

060

6 inch (150mm)

16

 

5311

25

8425

47

 

13186

75

 

20829

080

8 inch (200 mm)

27

 

8971

28

9487

53

 

14849

100

 

24687

100

10 inch (250 mm)

26

 

8637

49

13700

87

 

21443

155

 

34547

120

12 inch (300 mm)

36

 

12117

69

19220

91

 

22563

165

 

36660

140

14 inch (350 mm)

35

 

11693

67

18547

131

 

29030

235

 

47466

160

16 inch (400 mm)

55

 

15393

99

24417

189

 

38218

335

 

62026

200

20 inch (500 mm)

58

 

15989

114

25361

197

 

39696

375

 

64091

240

24 inch (600 mm)

92

 

22699

178

36006

304

 

56357

615

 

91094

INSTALLATION (WAFER SENSOR)

Gaskets

Alignment and Bolting

The following section should be used as a guide in the installation of the Rosemount 8711 Sensor. Refer to page 5-7for installation of the flange-type Rosemount 8705 and 8707 High-Signal sensor.

The sensor requires a gasket at each of its connections to adjacent devices or piping. The gasket material selected must be compatible with the process fluid and operating conditions. Metallic or spiral-wound gaskets can damage the liner. If the gaskets will be removed frequently, protect the liner ends. If grounding rings are used, a gasket is required on each side of the grounding ring.

1.On 11/2 - through 8-inch (40 through 200 mm) line sizes, place centering rings over each end of the sensor. The smaller line sizes, 0.15- through 1-inch (4 through 25 mm), do not require centering rings.

2.Insert studs for the bottom side of the sensor between the pipe flanges. Stud specifications are listed in Table 5-4. Using carbon steel bolts on smaller line sizes, 0.15- through 1-inch

(4 through 25 mm), rather than the required stainless steel bolts, will degrade performance.

5-10

Page 66
Image 66
Emerson Process Management 8732 manual Installation Wafer Sensor, Gaskets Alignment and Bolting, Size Code Line Size