Rosemount 8732

Reference Manual

00809-0100-4663, Rev BA January 2010

In Auto mode, OUT reflects the value and status quality of the output channels.

TROUBLESHOOTING

Refer to Table G-6to troubleshoot transducer block problems.

Table G-6. Troubleshooting

 

 

 

 

Symptom

Possible Causes

Corrective Action

 

Mode will not leave out of

Target mode not set

Set target mode to something other

 

service (OOS).

 

than OOS.

 

 

Resource block

The actual mode of the resource block

 

 

 

is OOS. See Appendix F: Resource

 

 

 

Block: and Section 3: Configuration.

 

PVor SV is BAD

Measurement

See Diagnostics, Table G-4.

 

 

 

Flow is above

 

 

 

SENSOR_RANGE.EU100.

 

PV or SV is UNCERTAIN

Measurement

Flow is above

 

 

 

PRIMARY_VALUE_RANGE.EU100.

G-6

Page 178
Image 178
Emerson Process Management 8732 Refer to Table G-6to troubleshoot transducer block problems, Table G-6. Troubleshooting