Reference Manual

00809-0100-4663, Rev BA January 2010

Rosemount 8732

FISCHER AND PORTER SENSORS

Model 10D1418 Sensor to Rosemount 8732 Transmitter

Connect coil drive and electrode cables as shown in Figure E-6.

Figure E-6. Wiring Diagram for

Fischer and Porter Sensor

Model 10D1418 and

Rosemount 8732

Electrode Connections

ROSEMOUNT 8732 TRANSMITTER

 

3

 

2

 

1

 

U1

 

U2

 

G

 

L1

 

L2

 

8

Coil Connections

6

 

7

 

5

 

Table E-7. Fischer and Porter

Model 10D1418 Sensor Wiring

Connections

Rosemount 8732

Fischer and Porter Model 10D1418 Sensors

1

 

 

 

L1

2

 

 

 

L2

 

 

 

 

 

 

Chassis Ground

 

 

 

 

 

17

 

3

18

 

1

19

 

2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.

E-9

Page 149
Image 149
Emerson Process Management manual Fischer and Porter Sensors, Model 10D1418 Sensor to Rosemount 8732 Transmitter