Rosemount 8732

Reference Manual

00809-0100-4663, Rev BA January 2010

TAYLOR SENSORS

Series 1100 Sensor to Rosemount 8732 Transmitter

Connect coil drive and electrode cables as shown in Figure E-19.

Figure E-19. Wiring Diagram for Taylor Series 1100 Sensors and Rosemount 8732

Table E-20. Taylor Series 1100 Sensor Wiring Connections

TAYLOR SERIES 1100

ROSEMOUNT 8732

SENSOR

TRANSMITTER

Electrode Connections

 

C

 

L

 

A

 

R

 

L N G 1 2 3 4

 

White

 

Black

 

Green

 

Coil Connections

 

Rosemount 8732

 

Taylor Series 1100 Sensors

1

 

 

 

Black

2

 

 

 

White

 

 

 

 

 

 

Green

 

 

 

 

 

17

 

 

S1 and S2

18

 

 

E1

19

 

 

E2

 

 

 

 

 

 

 

 

 

 

 

 

 

 

This is a pulsed DC magnetic flowmeter.

Do not connect AC power to the sensor or to terminals 1 and 2 of the transmitter, or replacement of the electronics board will be necessary.

E-22

Page 162
Image 162
Emerson Process Management manual Taylor Sensors, Series 1100 Sensor to Rosemount 8732 Transmitter