Reference Manual

00809-0100-4663, Rev BA January 2010

Rosemount 8732

Measurements

375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification

View the measured values taken during the meter verification process. These values are compared to the signature values to determine if the test passes or fails. Values are shown for the Coil Resistance, Coil Signature, and Electrode Resistance.

Coil Resistance

375Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, measurements

View the measured value for the coil resistance taken during the meter verification test.

Coil Signature

375Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, measurements

View the measured value for the coil signature taken during the meter verification test.

Electrode Resistance

375Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, measurements

View the measured value for the electrode resistance taken during the meter verification test.

Licensing

375 Transducer Block, Diagnostics, Advanced Diagnostics

If a diagnostic suite was not ordered initially, advanced diagnostics can be licensed in the field. Access the licensing information from this menu. For more details on licensing, see Appendix C: Diagnostics.

License Status

375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing

Determine if a diagnostics suite has been licensed, and if so, which diagnostics are available for activation.

License Key

375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing

A license key is required to activate diagnostics in the field if the diagnostic suite was not initially ordered. This menu allows for gathering of necessary data to generate a license key and also the ability to enter the license key once it has been received.

Device ID

375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing, License Key

This function displays the Device ID and Software Revision for the transmitter. Both of these pieces of information are required to generate a license key.

License Key

375 Transducer Block, Diagnostics, Advanced Diagnostics, Licensing, License Key

Allows you to enter a license key to activate a diagnostic suite.

4-7

Page 45
Image 45
Emerson Process Management 8732 manual Measurements, Licensing, License Status, License Key, Device ID