Emerson Process Management 8732 manual PTFE/ETFE liner Polyurethane liner, Class

Models: 8732

1 190
Download 190 pages 62.1 Kb
Page 64
Image 64

Rosemount 8732

Reference Manual

00809-0100-4663, Rev BA January 2010

NOTE

Do not bolt one side at a time. Tighten each side simultaneously. Example:

1.Snug left

2.Snug right

3.Tighten left

4.Tighten right

Do not snug and tighten the upstream side and then snug and tighten the downstream side. Failure to alternate between the upstream and downstream flanges when tightening bolts may result in liner damage.

Always check for leaks at the flanges after tightening the flange bolts. Failure to use the correct flange bolt tightening methods can result in severe damage. All sensors require a second torquing 24 hours after initial flange bolt tightening.

Table 5-1. Flange Bolt Torque Specifications for Rosemount 8705 and 8707 High-Signal Sensors

 

 

PTFE/ETFE liner

Polyurethane liner

 

 

 

 

 

 

 

 

Class 150

Class 300

Class 150

Class 300

Size Code

Line Size

(pound-feet)

(pound-feet)

(pound-feet)

(pound-feet)

 

 

 

 

 

 

005

1/2-inch (15 mm)

8

8

010

1 inch (25 mm)

8

12

015

11/2 inch (40 mm)

13

25

7

18

020

2 inch (50 mm)

19

17

14

11

030

3 inch (80 mm)

34

35

23

23

040

4 inch (100 mm)

26

50

17

32

060

6 inch (150mm)

45

50

30

37

080

8 inch (200 mm)

60

82

42

55

100

10 inch (250 mm)

55

80

40

70

120

12 inch (300 mm)

65

125

55

105

140

14 inch (350 mm)

85

110

70

95

160

16 inch (400 mm)

85

160

65

140

180

18 inch (450 mm)

120

170

95

150

200

20 inch (500 mm)

110

175

90

150

240

24 inch (600 mm)

165

280

140

250

300

30 inch (750 mm)

195

415

165

375

360

36 inch (900 mm)

280

575

245

525

See ”Safety Messages” on pages 5-1and 5-2for complete warning information.

5-8

Page 64
Image 64
Emerson Process Management 8732 manual PTFE/ETFE liner Polyurethane liner, Class