EP Trigger Level

375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits

Limits: 3 to 2000

Configure the threshold limit that the empty pipe value must exceed before the diagnostic alert activates. Default from the factory is set to 100. For more detail on this parameter see Appendix C: Diagnostics.

Advanced Diagnostics

375 Transducer Block, Diagnostics

EP Counts

375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits

Limits: 5 to 50

Configure the number of consecutive times that the empty pipe value must exceed the empty pipe trigger level before the diagnostic alert activates. Counts are taken at 1.5 second intervals. Default from the factory is set to 5. For more detail on this parameter see Appendix C: Diagnostics.

Electronics Temp Value

375 Transducer Block, Diagnostics, Basic Diagnostics

Electronics Temperature allows you to view the current value for the electronics temperature.

The advanced diagnostics menu contains information on all of the additional diagnostics and tests that are available in the 8732 transmitter if one of the diagnostics suite packages was ordered.

Rosemount offers two advanced diagnostic suites. Functionality under this menu will depend on which of these suites are ordered.

Advanced diagnostics suite option one (D01 option) contains advanced diagnostics for High Process Noise detection and Grounding and Wiring fault detection.

Advanced diagnostics suite option two (D02 option) contains advanced diagnostics for the 8714i Meter Verification. This diagnostic is used to verify the accuracy and performance of the magnetic flow meter installation.

8714i Meter Verification

375 Transducer Block, Diagnostics, Advanced Diagnostics

This diagnostic allows you to test and verify that the sensor, transmitter, or both are working within specifications. For more details on this diagnostic, see Appendix C: Diagnostics.

Run 8714i

375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification

Run the meter verification test to check the transmitter, sensor, or entire installation.

Full Meter Verification

Run the internal meter verification to check the entire installation, sensor and transmitter at the same time.

Transmitter Only

Run the internal meter verification to check the transmitter only.

Sensor Only

Run the internal meter verification to check the sensor only.