Reference Manual

00809-0100-4663, Rev BA January 2010

Rosemount 8732

Viewing the 8714i Meter Verification Results

Depending on the method used to view the results, they will be displayed in either a menu structure, as a method, or in the report format. When using the 375 Field Communicator, each individual component can be viewed as a menu item. In AMS, the calibration report is populated with the necessary data eliminating the need to manually complete the report found on

page C-16.

NOTE

When using AMS there are two possible methods that can be used to print the report. Method one involves taking a screen capture of the 8714i Report tab. Using Ctrl + Alt + PrntScrn will capture the active window and allow for pasting of the report directly into a word processing program.

Method two involves using the print feature within AMS while on the status screen. This will result in a printout of all of the information stored on the status tabs. Page two of the report will contain all of the necessary calibration verification result data.

The results are displayed in the following order:

Test Condition

Review the test condition that the 8714i Meter Verification test was performed under.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, 8714i Results, Test Condition

Context Menu, Device Diagnostics, 8714i Report

Test Criteria

Review the test criteria used to determine the results of the 8714i Meter Verification tests.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, 8714i Results, Test Criteria

Context Menu, Device Diagnostics, 8714i Report

8714i Result

Displays the overall result of the 8714i Meter Verification test as either a Pass or Fail.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, 8714i Results, 8714i Result

Context Menu, Device Diagnostics, 8714i Report

Simulated Velocity

Displays the simulated velocity used to verify the transmitter calibration.

375

AMS Tab

Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification, 8714i Results, Simulated Vel

Context Menu, Device Diagnostics, 8714i Report

C-11

Page 131
Image 131
Emerson Process Management 8732 manual Viewing the 8714i Meter Verification Results, Test Condition