Inlet Systems

Packed column inlet

Assuming the system to be leak•free(and if total flow is < 200 ml/min), after setting the desired column flow rate, total flow through the system should be noted via the EFS.

The original column flow rate is reestablished simply by adjusting the mass flow controller so the original total system flow rate value is displayed again.

Septum-purged packed column inlet

The septum•purgedpacked column inlet may be used with HP Series 530 ¿ capillary columns, metal packed or glass packed columns. Additionally, on•columninjection is possible with 1/4•inchglass packed columns.

The septum•purgedinlet provides superior reproducibility and quantitative accuracy at high inlet temperatures (250 to 400^C).

Figure 6-3

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

External

Internal

 

 

 

 

 

 

 

 

 

 

 

Septum Purge

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Packed Column

Trap(s)

 

Plumbing

Plumbing

Pressure

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gauge

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

To

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Carrier

 

 

 

Mass

 

Electronic

 

 

 

 

Detector

 

 

 

Flow

 

Flow Sensor

Gas

 

 

 

 

 

 

 

 

 

 

Controller

 

(optional)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Forward

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pressure Regulator

 

 

 

 

Septum

 

Out

 

In

Column

 

 

 

 

 

 

 

Purge

Fixed

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Vent

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Restrictor

 

 

 

 

 

 

 

 

 

 

 

 

Flow Diagram, Septum-Purged Packed Column Inlet