Detector Systems

Nitrogen-phosphorus detector (NPD)

Residual silanizing reagents from derivatization, and/or bleed from silicone columns, may coat the active element with silicon dioxide. This decreases ionization efficiency, reducing sensitivity.

If silanizing is necessary, remove excess reagent before injection. Silicone columns should be well conditioned and loaded less than 5%.

Active element lifetime

Lifetime of the active element is reduced by the silicon dioxide coating, described above, and by irreversible loss of rubidium salt.

Rubidium loss is caused by overheating the active element, particularly if element power is on when gas flows, particularly carrier, are interrupted. The detector must be turned off or element power reduced to zero when changing columns, replacing septa, and/or replacing gas cylinders. Power to the element with gas flow off can destroy an element within a few minutes.

Humidity also affects element lifetime adversely: keep the detector warm (100 to 150 C) when not in use. A collector removed from an NPD for an extended period of time, or spare collectors, should be stored in a desiccator.

Element lifetime is extended by using the lowest element power possible, consistent with maintaining sufficient detector sensitivity and selectivity for the particular analyses.

Generally, sensitivity and selectivity toward nitrogen decreases first as the element ages. Phosphorus response is affected less.

It is good practice to have a spare collector available for replacement purposes. A recoating kit is available (Part No. 5080•8872)to rejuvenate the active element in an old collector.