Test Sample Chromatograms
Test sample chromatograms
Figure
HP 5890 Test Sample Operating
Conditions
Detector Type NPD w/MUG Temp 220 DEGREES C
Inlet Type Split only or split/splitless
Temp 200 DEGREES C
Operating Mode Split(Purge on)
Purge Time On | 0 |
| min | |
Purge Time Off | 0 |
| min | |
Oven Isothermal |
|
|
| |
Init Temp | 180 | DEGREES C | ||
Init Time | 5.0 | min | ||
Ramp |
|
|
|
|
Rate |
| 0 |
|
|
Fin Temp |
|
|
|
|
Fin Time |
|
|
|
|
Range 0 |
|
|
|
|
COLUMN: |
|
|
|
|
Part No. | 19095Z(#121) | |||
Dimensions |
| 530¿ | MID; 10 M | |
Sta Phase |
|
|
|
|
FLOW RATES |
|
|
|
Carrier (He) | 15 +/• 1 | ml/min | |
Hydrogen |
| 3.5 +/• 0.1 | ml/min |
Air | 100•120 +/• 10 | ml/min | |
Makeup (He) | 10+/•1 | ml/min | |
Split Vent |
| 80+/•5 | ml/min |
Septum Purge | 5+/•1 | ml/min |
SAMPLE:
Type
START
1.35 AZOBENZENE
2.37 C18
3.22 MALATHION
ST