Preventive Maintenance
Electron capture detector (ECD)
Packed column:
1.Close the anode purge on/off valve.
2.Remove the column from the detector; install in its place an empty glass column.
3.Establish normal carrier gas flow rate (20 to 30 ml/min); set oven temperature to 250^C.
4.Open the anode purge on/off valve.
5.Heat the ECD to 350^C. Allow thermal cleaning to continue for 3 to 12 hours. If the displayed value does not drop below 60 (600Hz), indicating a high background signal, it is recommended that the detector be returned to
Capillary column:
1.Close the makeup gas and anode purge on/off valve.
2.Remove the column from the makeup gas detector.
3.Install a paper clip or similar gauge wire in the same manner as a capillary column.
4.Open the makeup gas on/off valve and establish a makeup gas flow rate of 50 to 60 ml/min.
5.Open the anode purge on/off valve.
6.Set the oven temperature to 250^C.
7.Heat the ECD to 350^C. Allow thermal cleaning to continue for 3 to 12 hours. If the displayed value does not drop below 60 (600Hz), indicating a high background signal, it is recommended that the detector be returned to Hewlett•Packardfor cleaning.
191