Temperature Control

Oven status

In isothermal operation ( RATE = 0 ), if INIT TIME is set equal to 0 (zero), the HP 5890 internally sets run time to the maximum, 650 minutes.

A is included in key sequences defining parameters for a second ramp; B is included in key sequences defining parameters for a third ramp.

In isothermal operation, and in one•or two•ramptemperature programs, rate for the next ramp must be set to 0 (zero) to prevent further programming.

In oven temperature programming, once any one of the five temperature

programming functions ( INIT TEMP

,

INIT TIME , RATE , FINAL TEMP , and

FINAL TIME ) is displayed, pressing

ENTER

, without entering a new

setpoint value, rolls the display successively through the entire group, (including A and B displays for second and third ramps).

This is an efficient way in which to review and, if necessary, change oven temperature program setpoints.

Oven status

During a temperature•programmedrun, the LED OVEN display provides indication of oven status at any given time:

CIsothermal Run: Assuming INIT TIME is assigned a value greater than 0, and that RATE = 0, only the INITIAL TIME LED is lit. It remains lit throughout the run.

CSingle•Ramp Temperature Program: The three LEDs, INITIAL TIME, RATE, and FINAL TIME, successively light to indicate position in the temperature program.

CMultiple•Ramp Temperature Program: The three LEDs, INITIAL TIME, RATE, and FINAL TIME, successively light to indicate position in the first temperature program ramp.

Then RATE and FINAL TIME LEDs light alternately as the program proceeds through the second (and third) ramp(s).

50