Detector Systems

Nitrogen-phosphorus detector (NPD)

Other gas flow effects of too high flow rates of the hydrogen may allow a true flame to exist around the active element. This would overheat the active element severely and destroy the specific response. Too low flow rates of air tend to quench the background response of the active element, and this results in a re•equilibrationtime that is too long to establish proper background response (negative solvent peaks killing the active element).

 

Figure 7-4

Internal

 

 

 

 

 

 

 

 

 

 

External

 

 

 

 

 

 

 

 

 

 

Plumbing

Plumbing

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Active Element

 

 

 

 

 

 

 

 

 

 

 

Final Pressure Regulators

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Power

 

 

 

 

 

 

 

 

 

and Gauges

 

 

 

 

 

 

 

 

 

 

 

 

 

Detector

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Traps

 

 

 

 

 

 

 

 

 

 

 

Power to

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Active Element

 

 

 

 

Compressed

 

 

Flow Manifold Block

 

 

 

 

 

 

 

 

Air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AIR

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On/Off

 

 

 

 

Air

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Hydrogen

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gas

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

HYDROGEN

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On/OffHydrogen

 

 

 

 

 

 

 

 

 

 

From

 

 

 

 

 

 

 

 

 

 

 

 

 

(Plus Makeup

 

 

 

 

 

 

 

 

 

 

 

 

 

Inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

Gas)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Capillary

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Makeup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Gas

 

 

 

 

 

 

 

MAKEUP GAS

 

Column

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On/Off

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow Diagram, Nitrogen-Phosphorus Detector (NPD)

 

 

 

131