Test Sample Chromatograms

Test sample chromatograms

Figure 10-8.

HP 5890 Test Sample Operating

Conditions

Detector Type ECDw/MUG

FLOW RATES

 

 

Temp 300 DEGREES C

Carrier (He)

15 +/• 1

ml/min

Inlet Type Split only or

Hydrogen

N/A

ml/min

splitless

 

Air

N/A

ml/min

Temp 200 DEGREES C

 

Makeup (N2)

60 +/• 2

ml/min

Operating Mode Split(Purge on)

Split Vent

60 +/• 2

ml/min

Purge Time On

0

 

min

Septum Purge

5 +/• 1

ml/min

Purge Time Off

0

 

min

 

 

 

Oven Temp Isothermal

 

 

 

 

Init Temp

170

DEGREES C

 

 

 

Init Time

N/A

min

SAMPLE:

 

 

Ramp

 

 

 

 

 

 

Rate

 

0

 

 

Type ECD Sample

 

Fin Temp

 

 

 

 

Inj Volume

1 ¿l

 

Fin Time

 

 

 

 

Part No. 18713•60040

 

Range 0

 

 

 

 

Composition 33 pg/¿l (0.033

 

 

 

 

 

ppm (W/V)) each: lindane

 

COLUMN:

 

 

 

 

and aldrin in isooctane

 

Part No.

19095S(#100)

 

 

 

Dimensions

 

530¿ MID; 5 m

 

 

 

Sta Phase

Methyl Silicone

 

 

 

.37

.43

1.13 LINDANE

2.64ALDRIN

ST

ECD-Split Mode Capillary Inlet

232