Inlet Systems
Split/splitless capillary inlet
Table
Various Capillary Column Bores and Lengths
Nominal |
| Nominal Length (m) |
|
|
|
|
|
| |
ID (mm) | 12 | 25 | 50 |
|
|
|
|
|
|
0.20 | 135 | 223 | 347 |
|
0.32 | 45 | 82 | 137 |
|
0.53 | 11 | 23 | 42 |
|
It must be emphasized that values in this table are recommended as starting points only!
Values listed are independent of carrier gas used.
It is important to note that flow settings made for one particular column are not necessarily correct for any other column, or for every application. Column flow rate should be verified, and adjustments made as necessary, whenever the column or other parts of the flow system are changed.
Split sampling
WARNING In performing split sampling, when using hazardous chemicals, and/or H2 carrier gas, effluent from the split vent should be vented to a fume hood or appropriate chemical trap.
111