Signal Output

Test signal output

Test signal output

A test chromatogram, consisting of three peaks, is permanently stored in the HP 5890. Each peak is approximately 1/10 the height of the previous peak, with the first (tallest) peak having a height value of about 125 mV at RANGE 2!( ) = 0 (+ 1 V analog output); half•heightwidth of this peak is about 0.13 minutes.

Figure 5-5

HP 5890:

RANGE = 0 2 ” ( )

LIST: LIST

PEAK CAPACITY: 1159

ZERO = 3, •1.5 ATT 2! = 7

CHT SP = 1.3 PK WD = 3.13 THRSH = 3 AR REJ = 3

START

ATTN

2 ” ( )

= 0

.25

1 V Analog Output

.75

1.25

STOP

RUN # 8

AREA%

 

 

 

 

RT

AREA

TYPE

AR/HT

AREA%

3.25

7939433

BB

3.135

93.397

3.75

793533

PB

3.135

9.335

1.25

79394

BB

3.134

3.898

TOTAL AREA

=

8812133

MUL FACTOR

=

1.3333E+33

Typical HP 5890 Test Chromatogram (using an HP 3390A Reporting Integrator)