Test Sample Chromatograms

Test sample chromatograms

Figure 10-5.

HP 5890 Test Sample Operating

Conditions

Detector Type

FIDw/MUG

 

FLOW RATES

 

 

 

Temp 250 DEGREES C

 

Carrier (He)

15 +/• 1

ml/min

Inlet Type Ded On-Col Cap

 

Hydrogen

30 +/• 1

ml/min

Oven Track

 

 

On

 

 

Air

400 +/• 20

ml/min

 

Temp N/A

 

C

Makeup (N2)

20 +/• 1

ml/min

Operating Mode

 

N/A

 

Split Vent

N/A

ml/min

Purge Time On

 

N/A

min

Septum Purge

5 +/• 1

ml/min

Purge Time Off

 

N/A

min

 

 

 

Oven Temp Temp Programmed (1

 

 

 

ramp)

 

 

 

 

 

 

 

 

 

 

Init Temp

60

DEGREES C

 

 

 

Init Time

0.5

min

 

 

 

 

Ramp

 

 

 

 

 

 

 

SAMPLE:

 

 

Rate

 

 

20

 

 

 

 

 

 

 

 

 

 

Type FID Sample

 

Fin Temp

 

 

185

 

 

 

 

Fin Time

 

 

1

 

 

 

Inj Volume

1 ¿l

 

Flow Param (EPP)

 

 

Part No. 18710•60170

 

 

 

Composition 0.03%(V/V)each:

Constant Flow

Off

 

 

C14,C15, and C16 normal paraffin

Range

4

 

 

 

 

 

 

 

 

 

 

 

 

hydrocarbons in hexane.

 

 

 

 

 

 

 

 

 

 

COLUMN:

 

 

 

 

 

 

 

 

 

 

Part No.

 

19095(#100)

 

 

 

 

Dimensions

 

 

530¿ MID; 5 M

 

 

 

Sta Phase

Methyl Silicone

 

.112

 

 

 

 

 

 

 

 

START

 

 

 

 

 

 

 

 

 

 

.331

.267

4.743 C14

5.398 C15

6.322 C16

FID-On-Column Capillary Inlet

229