Preventive Maintenance

Electron capture detector (ECD)

Electron capture detector (ECD)

Frequency test

Note: For high sensitivity operation, and starting from a cold system, 24 hours may be necessary before baseline is completely stabilized. Use low•bleedsepta and condition a new septum prior to use in an unused inlet for several hours with 1 to 5 ml/min carrier flow rate.

Using normal operating conditions, and allowing at least two hours after the last sample was injected, enter the following:

DET A ( or B ) ON

SIG 1 ( or SIG 2 ) ( A or B )

If the displayed value exceeds 100 (1000 Hz), indicating a high background signal, investigate the following possibilities:

CContaminated carrier gas trap(s) and/or supply

CInsufficient column conditioning

CContaminated detector

CColumn, inlet, and/or septum bleed

CLeaks

CAnode insulation leakage

Carrier gas evaluation

Because of its very high sensitivity, the ECD is particularly prone to contamination problems, including contaminants entering the system via the carrier gas source.

The following procedure should be performed whenever a new carrier gas source is installed:

1.With the instrument on and operating normally, cool the oven to ambient, turn off the detector, turn off carrier flow to the detector, and

188