Detector Systems

Electron capture detector (ECD)

A very clean system may produce a value below the low end of 10 (100 Hz). To correct this condition, an adjustment is made to the present potentiometer located on the ECD electronics board.

ECD Potentiometer SwitchUp: Adj

Down: Fixed

ECD Potentiometer Adjustment

ECD Potentiometer Switch and Adjustment

1.Remove the right side of the panel.

2.Flip the switch up to the Adj. position.

3.Using a small screwdriver, turn the potentiometer clockwise until the background level is at or slightly above 20 (200 Hz, low reading for trace analysis).

4.Replace the right side of the panel.