Detector Systems

Thermal conductivity detector (TCD)

Because of its exceptionally high thermal conductivity and chemical inertness, He is the recommended carrier gas: it gives large thermal conductivity differences with all compounds except H (considerations necessary in H analyses are discussed later). With He as carrier, the TCD exhibits universal response. For propane, the sensitivity limit is about 400 picograms/ml of He carrier gas.

The TCD exists in two configurations: either its exhaust vent tube exits at the top of the detector or the exhaust vent tube returns to the inside of the oven for connection to an FID or other device.

 

 

Figure 7-8

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

External

Internal

 

 

 

 

 

 

 

 

 

 

Detector

 

 

 

Plumbing

Plumbing

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Final Pressure Regulators

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

and Gauges

 

 

 

 

 

 

 

 

 

 

 

Filament

 

 

 

 

 

 

 

 

 

 

 

Electronics

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Traps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Solenoid Switching

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow Manifold BlockValve

 

 

 

 

 

Reference

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

REFERENCE

 

 

 

 

 

 

 

 

 

Gas

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On/Off Reference

 

 

 

From

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Makeup Gas

Inlet

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Capillary

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MAKEUP GAS

 

 

 

 

 

Column

 

 

 

Makeup

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

On/Off

 

 

 

Gas

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Makeup Gas Adapter:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Capillary Columns Only

 

 

 

Flow Diagram, Thermal Conductivity Detector (TCD)

 

 