Electronic Flow Sensing
Electronic flow sensor (EFS) calibration
3.Locate the EFS module and note its labelling: CHANNEL A/ CHANNEL B, IN/OUT. For the channel being calibrated, locate and disconnect its OUT fitting; use two wrenches in opposition to prevent twisting the tubes.
Figure
Outlet Line, Channel B | Outlet Line, Channel |
| A |
EFS Module
Detail, Electronic Flow Sensor (EFS) Module
4.Install the EFS flow•measuringadapter (Part No. 05890•80620)into the female OUT fitting to the EFS module. Connect a bubble flow meter to the adapter.
Allow ample time (up to 1/2•hour)for residual gas within connected plumbing to bleed off. Verify that absolutely no flow is observed at the connected bubble flow meter.
62