Signal Output

Test signal output

To access this function, the following key sequence is entered:

SIG 1

(or

SIG 2

)

9

ENTER

 

 

 

Test plot mode is confirmed by the display

 

SIGNAL 1 (or 2) TEST

PLOT. Pressing SIG 1 (or

SIG 2

) a second time displays the current

signal level value (which is

0.0

initially). This permits monitoring the

output signal.

 

 

 

 

 

The chromatogram is initiated by pressing START. The chromatogram continues to cycle until STOP is pressed. Each cycle is about 1•1/2 minutes in length.

Note:

CSetpoint values assigned to RANGE 2!( ) and ATTN 2!( ) affect level of the test signal output, as do equivalent functions on the receiving

device.

CAn oven temperature program (e.g., a setpoint value for RATE other than 0) must be set up at the HP 5890 for the test plot to function.

The test chromatogram is useful as a troubleshooting aid in deciding whether a lost or noisy signal observed at a connected integrating or chart recording device is due to a chromatographic problem (lost sample due to leaks, noise due to a dirty detector, etc.), versus problems either with the integrating/recording device itself, or in its connecting cables.

If the test chromatogram does not exhibit any problems at the integrating/recording device, a chromatographic problem is likely to exist; if the test chromatogram exhibits noise, or does not appear at all, the problem is not likely to be chromatographic.