Temperature Control
Cryogenic
Cryogenic
Liquid N1 or liquid CO1 cryogenic options are for operation at temperatures less than about 7^C above ambient. This is done through operation of a valve which opens when coolant is demanded and closes when the setpoint temperature is reached.
When you press gold CRYO PARAM CRYO PARAM CRYO PARAM you scroll
through a series of displays for choosing cryogenic options. These options include CRYO for operation during the entire run, CRYO BLAST, for very fast cool down between runs, and AMBIENT to regulate on and off times to optimize coolant use.
Figure
ACTUAL | SETPOINT | For operation during runs at | |
CRYO | ON | ||
subambient temperatures. | |||
ACTUAL | SETPOINT | ||
|
CRYO BLAST | ON |
ACTUAL | SETPOINT |
|
|
AMBIENT | 25 |
For very fast cool down between runs.
Lets you regulate Cryo and Cryo Blast on and off times to optimize coolant use. Default is 25^C and need not be changed unless
ambient differs by 10^.
Cryogenic options
The following figures show the oven temperature profile for a typical run, showing the on and off times for CRYO and CRYO BLAST.
47