Test Sample Chromatograms
Test sample chromatograms
Figure
HP 5890 Test Sample Operating
Conditions
Detector Type TCD(or TCDw/MUG) | FLOW RATES |
|
| ||||||||
| Temp 300 DEGREES C | Carrier (He) | 20 +/• 1 | ml/min | |||||||
Inlet Type PACKED (OR PURGED | Hydrogen | N/A | ml/min | ||||||||
|
| PACKED). |
| Air |
|
| N/A | ml/min | |||
| Temp 250 DEGREES C | Makeup (He) | 1•2 | ml/min | |||||||
Operating Mode |
| N/A |
| Split Vent | N/A | ml/min | |||||
Purge Time On |
| N/A | min | Septum Purge | 1•2 | ml/min | |||||
Purge Time Off |
| N/A | min | Reference (He) 30+/•1 | ml/min | ||||||
Oven Temp Programmed (1 ramp) |
|
|
|
|
| ||||||
Init Temp | 110 | DEGREES C |
|
|
|
|
| ||||
Init Time | 1 | min |
|
|
|
|
| ||||
Ramp |
|
|
|
|
|
| SAMPLE: |
|
| ||
Rate |
|
| 15 |
|
|
|
| ||||
Fin Temp |
|
| 150 |
|
| Type |
| FID Sample |
| ||
Fin Time |
| 1 |
|
| Inj Volume | 3 ¿l |
| ||||
Range | 0 |
|
|
|
|
| Part No. 18710•60170 |
| |||
|
|
|
|
|
|
| Composition 0.03%(V/V)each |
| |||
COLUMN: |
|
|
|
|
|
| C14,C15,C16 normal paraffin | ||||
Part No. |
| 19095S(#100) | hydrocarbons in hexane |
| |||||||
Dimensions |
|
| 530¿ MID; 5 M |
|
|
|
|
| |||
Sta Phase | Methyl Silicone |
|
|
|
|
| |||||
|
| START |
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
| .25 |
|
|
|
|
|
|
|
| 2.42 | C14 |
|
| |
|
|
|
|
|
|
| 3.38 |
| C15 |
|
|
|
|
|
|
|
|
| 3.76 | C16 |
|
| |
|
|
|
|
| ST |
|
|
|
|
|
Thermal Conductivity Detector (TCD)
228