Test Sample Chromatograms

Test sample chromatograms

Figure 10-4.

HP 5890 Test Sample Operating

Conditions

Detector Type TCD(or TCDw/MUG)

FLOW RATES

 

 

 

Temp 300 DEGREES C

Carrier (He)

20 +/• 1

ml/min

Inlet Type PACKED (OR PURGED

Hydrogen

N/A

ml/min

 

 

PACKED).

 

Air

 

 

N/A

ml/min

 

Temp 250 DEGREES C

Makeup (He)

1•2

ml/min

Operating Mode

 

N/A

 

Split Vent

N/A

ml/min

Purge Time On

 

N/A

min

Septum Purge

1•2

ml/min

Purge Time Off

 

N/A

min

Reference (He) 30+/•1

ml/min

Oven Temp Programmed (1 ramp)

 

 

 

 

 

Init Temp

110

DEGREES C

 

 

 

 

 

Init Time

1

min

 

 

 

 

 

Ramp

 

 

 

 

 

 

SAMPLE:

 

 

Rate

 

 

15

 

 

 

 

Fin Temp

 

 

150

 

 

Type

 

FID Sample

 

Fin Time

 

1

 

 

Inj Volume

3 ¿l

 

Range

0

 

 

 

 

 

Part No. 18710•60170

 

 

 

 

 

 

 

 

Composition 0.03%(V/V)each

 

COLUMN:

 

 

 

 

 

 

C14,C15,C16 normal paraffin

Part No.

 

19095S(#100)

hydrocarbons in hexane

 

Dimensions

 

 

530¿ MID; 5 M

 

 

 

 

 

Sta Phase

Methyl Silicone

 

 

 

 

 

 

 

START

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.25

 

 

 

 

 

 

 

 

2.42

C14

 

 

 

 

 

 

 

 

 

3.38

 

C15

 

 

 

 

 

 

 

 

 

3.76

C16

 

 

 

 

 

 

 

ST

 

 

 

 

 

Thermal Conductivity Detector (TCD)

228