Signal Output
Figure
|
| ACTUAL | SETPOINT |
|
|
|
|
HPIL | LOOPBACK | TEST |
|
|
| ACTUAL | SETPOINT |
|
|
| |
PASSED SELF | TEST |
| |
|
| ACTUAL | SETPOINT |
|
|
|
|
FAILED | SELF | TEST |
|
HPIL LOOPBACK TEST Displays
The message PASSED SELF TEST indicates INET, at least with respect to the HP 5890, is performing satisfactorily. If FAILED
Note: The loopback test may be used to check for continuity in an INET cable; an open cable causes test failure. Verify that the cable is at fault (rather than the HP 5890) by testing a second cable.
If an intermittent cable problem is suspected, the test may be repeated as necessary while flexing the cable (particularly at its plugs). An ohmmeter should also be used to test for problems; it is a reliable method for testing continuity.
91