Keyboard and Displays
Loading default setpoints
With the keyboard locked, Figure 2•6shows the display occurring if a setpoint entry is attempted:
Figure
ACTUAL SETPOINT
KEYBOARD LOCKED
KEYBOARD LOCKED Message Display
If the HP 5890 keyboard is locked while the instrument is under INET control, a setpoint file may be loaded into HP 5890 memory from the controller, but the loaded setpoints cannot then be edited at the HP 5890 keyboard until it is unlocked.
After locking or unlocking the keyboard, return to operation by pressing any function key (e.g., OVEN TEMP ).
Loading default setpoints
This function permits resetting HP 5890 operating setpoints to a standard set of values. Only the following information is retained:
CCalibration constants for oven temperature control
CAll gas flow rate monitoring information (gas types and calibration constants)
CColumn compensation data, including detector assignments
Note: Since user•definedsetpoints are lost in the process, any critical HP 5890 setpoints should be recorded for later reentry before proceeding.
Through the keyboard, select CALIB AND TEST mode, function 6:
CLEAR . 6
39