Preventive Maintenance
Thermal conductivity detector (TCD)
Radioactivity leak test (wipe test)
ECDs must be tested for radioactive leakage at least every six months. Records of tests and results must be maintained for possible inspection by the Nuclear Regulatory Commission and/or responsible state agency. More frequent tests may be conducted when necessary.
The procedure used is the wipe test. A wipe test kit (Part No. 18713•60050)is supplied with each new ECD. Its contents are listed in Table 8•2:
Table
Item | Description | Part No. | Qty |
|
|
|
|
1 | Envelope | 4 | |
2 | Envelope | 4 | |
3 | Information Card | 12 | |
4 | Filter Paper, Circular | 12 | |
5 | Service Note | 1 | |
6 | Plastic Bag | 12 | |
|
|
|
|
Thermal conductivity detector (TCD)
Cleaning
Cleaning by bake•out of the detector is performed only after verification that carrier supply gas and flow system components, are leak•and contaminant•free.
1.Turn off the detector!
DET | A ( or B ) OFF |
2. Remove the column from the detector; cap the detector column fitting.
192