Test Sample Chromatograms

Test sample chromatograms

Figure 10-9.

HP 5890 Test Sample Operating

Conditions

Detector Type

TCDw/MUG

FLOW RATES

 

 

 

 

Temp 300 DEGREES C

Carrier (He)

15 +/• 1

 

ml/min

Inlet Type Split only or

Hydrogen

N/A

 

ml/min

 

split/splitless

Air

 

N/A4

 

ml/min

Temp

250 DEG C

 

 

 

 

Makeup (He)

10 +/• 1

 

ml/min

Operating Mode Split(Purge on)

Split Vent

200 +/• 10

 

ml/min

Purge Time On

0

 

min

Septum Purge

5 +/• 1

 

ml/min

Purge Time Off

0

 

min

Reference (He)37 +/• 1

 

ml/min

Oven Temp Programmed

 

 

 

 

 

 

Init Temp

100

DEGREES C

 

 

 

 

 

Init Time

1

min

 

 

 

 

 

Ramp

 

 

 

 

 

 

 

 

 

 

Rate

 

 

10

 

 

SAMPLE:

 

 

 

 

 

 

 

 

 

 

Fin Temp

 

 

150

 

 

Type

Capillary

 

 

Fin Time

 

 

2

 

 

Inj Volume

1 ¿l

 

 

 

 

 

 

 

 

Range

0

 

 

 

 

 

Part No. 18740•60900

 

 

 

 

 

 

 

 

 

Composition See Split/FID

 

COLUMN:

 

 

 

 

 

 

 

 

 

 

Part No.

19095Z(121)

 

 

 

 

 

Dimensions

 

 

530¿ MID; 5 M

 

 

 

 

 

Sta Phase

Methyl Silicone

 

 

 

 

 

 

 

 

 

 

 

START

.11

 

.15

.42

 

 

 

 

 

 

 

 

.25

 

 

 

 

 

 

 

 

 

 

 

 

 

C9

TCD-Split Mode Capillary Inlet

3.64

4.78

ST

3.85

4.45

6.48

C16

233