Test Sample Chromatograms

Test sample chromatograms

Figure 10-15.

HP 5890 Test Sample Operating

Conditions

Detector

Type FPD

 

 

 

Temp 200 DEGREES C

 

Inlet Type DED ON-COL CAP

 

Oven Track

 

 

On

 

 

Temp N/A

DEGREES C

 

Operating

Mode

 

N/A

 

Purge Time On

 

N/A

min

Purge Time Off

 

N/A

min

Oven Temp Programmed (1 ramp)

 

Init Temp

90

DEGREES C

Init Time

1.0

min

 

Ramp

 

(1)

(2)

 

Rate

 

20

10

 

Fin Temp

110

170

 

Fin Time

0

3

 

Flow Param (EPP)

 

 

Constant Flow

Off

 

Range 5

 

 

 

 

COLUMN:

 

 

 

 

Part No.

 

19095S•100

 

Dimensions

 

5 M Length, 530 ¿

ID

Sta Phase Methyl Silicone

FLOW RATES (FPD AIR MODE FLOWS)

 

Carrier (N2)

 

20 +/• 1

ml/min

Hydrogen

 

75 +/• 2

ml/min

Air

 

 

100 +/• 3

ml/min

Makeup (N2)

 

0

ml/min

Split Vent

 

N/A

ml/min

Septum Purge

 

5 +/• 1

ml/min

SAMPLE:

 

 

 

 

Type

FPD Sample

 

Inj Volume

1

¿l

 

Part No. 19395•60580

 

COMPOSITION: 20 ng/¿L

 

(20.0 ppm W/V) each of

 

1-dodecanethiol

and

 

tributylphosphate in isooctane

START .13

START

.13

 

 

 

 

Sulfer

.49

Phosphorus

 

 

1.91

Filter

1.45

Filter

 

 

 

4.06 DODECANETHIOL

 

4.08

 

4.57

3.35

TRIBUTYLPHOS-

5.37

 

 

PHATE

IF

IF

 

FPD On-column Inlet

239