Signal Output

File compatibility with data handling devices

How to convert HP 339X Integrator workfiles from 5890A to SERIES II mode:

1.Turn the GC off.

2.Follow the previous instructions to set the GC for 5890A mode (use proper grounding).

3.Download the workfile from the integrator.

4.Turn GC power off.

5.Remove the P15 jumper. (Now the GC is in SERIES II mode.)

6.Turn GC power on.

7.Add SERIES II setpoints (time table, etc).

8.Store the workfile at the integrator.

Note: Using STORE on the HP 5890 SERIES II, up to 4 integrator workfiles can be converted with each jumper change sequence.