Columns and Fittings

Liners/adapters and inserts, general

Table 1-2. Hardware and Recommended Fittings for Packed Column Installation

 

 

Packed Columns

 

 

1/8-inch Metal

1/4-inch Metal

1/4-inch Glass

 

 

 

 

Recommended

1/8-inch

1/4-inch

1/4-inch swage-

Column Fittings

swage-type nut

swage-type nut

type nut and

 

and ferrules3

and ferrules3

graphic ferrule or

 

 

 

silicone O-ring(s)

Packed Column

19243-805101

19243-805201

None

Inlet Liners

or

or

 

 

19243-805301

19243-805401

 

 

(requires glass insert)

(requires glass insert)

 

FID/NPD2

19231-805211

19231-805301

None

Liners/Adapters

 

 

 

TCD Liners/Adapters

None

19302-800201

19302-800201

 

 

 

(may require

 

 

 

altering the

 

 

 

column)

ECD Liner/Adapters

19301-805301

None

None

1Use 1/4-inch swage-type nut and Vespel or graphite ferrule to install liner/adapter.

2See Chapter 8 for details regarding jet exchange (if necessary).

3See information later in this chapter regarding proper installation of swage-type fittings on packed metal columns.