Temperature Control
Oven temperature calibration
3.CALIB DELTA is displayed until ENTER is pressed; then oven temperature recalibration occurs. Note that, after calibration, the
displayed oven temperature value should match closely the measured value.
Any delta value within the range
Assuming the battery protecting HP 5890 memory is operational, a new calibration constant remains in effect even if the instrument is switched off, or disconnected from its power source, or if power fails.