Preventive Maintenance
Liner and/or insert care
Repacking a split insert
Since, for a split insert, its packing material is discarded in cleaning, the insert must be repacked.
Note: Repacking with small•diameterglass beads is not recommended: they are usually contaminated with metal filings due to sieving procedures used. If they must be used, thorough cleaning (chemical and physical) is required.
CUse a fresh, small amount of a conventional coated packing such as 2% OV•1on 100/120 mesh, Chromosorb W•HP.
CPacking is held in place between plugs of silanized glass wool (Part No. 8500•1572).
Figure
Glass | 35 mm |
Wool |
Packing
Packing Requirements, Split Insert
CTo ensure compatibility with standard 2•inchsyringe needles, there should be at least 35 mm between the top of the insert and the top of the packed section.
170