Detector Systems

Thermal conductivity detector (TCD)

Optimizing performance

The following sections aid in choosing operating parameters to obtain optimal TCD performance.

Temperature

TCD sensitivity increases as the temperature difference between the detector filament (automatically set) and the surrounding detector body (chosen detector zone temperature) increases.

To minimize changes in detector sensitivity with different operating temperatures, filament temperature is maintained at a relatively constant difference above detector operating temperature. Filament resistance (proportional to temperature) is monitored, and power is adjusted as needed to maintain the filament at constant temperature.

Figure 7-9

500

 

 

400

Filament Temperature

50 C

 

 

Filament 300

Temp

( C)

200

T

Block Temperature

 

 

 

 

 

 

 

100

135 C

 

 

 

 

 

 

 

0

 

 

 

 

0

100

200

30

400

 

Detector Temperature (Body 0

 

 

Temp)

C

 

 

Filament Temperature versus Detector Block Temperature

146