Inlet Systems
Split/splitless capillary inlet
Figure
Area
Counts
+1.2% Deviation
~20 ppm
+ 1.2% Deviation
~10 ppm
Solvent: Isooctane
Column: 16.5 m x 0.25 mm
80^C (0.5 min) – 170 C @ 15 /min Sample Size: 1.3 ml (Manual Injection) Injection Rate: 1 ml/sec
Purge Flows: 5 ml/min, Septum: 60 ml/min, Inle
10 | 20 | 30 | 40 | 50 | 60 |
Purge Activation Time, Sec
Effect of Inlet Purge Activation Time on Area Counts
Noting Figure 6•10,waiting too long does not increase component peak areas, but does increase interference by the solvent tail. Purging too early risks venting light components, not allowing sufficient time for heavier components to enter the column, and/or not having sufficient solvent enter the column to ensure good reconcentration.
A recommended procedure is to analyze a known standard using conditions identical to those to be used for later sample analyses. Ideally, the standard should be representative (in both components chosen and their respective concentrations) of unknown samples to be analyzed.
Perform a series of analyses with increasingly delayed insert purge times: the optimal time is where maximum area counts are obtained with least interference by the solvent.
119