Preventive Maintenance

Conditioning chemical traps

Conditioning chemical traps

Remove the trap from its installed location and attach it to a clean, dry gas source (helium or nitrogen). Attach the 1/8•inchend (male) of the chemical trap assembly to the reconditioning gas source using a graphite or a graphitized Vespel ferrule (Part No. 0100-1 107) and swage nut (Part No. 0100-0058), as metal type ferrules will damage the sealing surface of the 1/8•inchadapter. Remove the O•ring(Part No. 5180•4181),as the conditioning temperature will destroy it. Set oven temperature to 300^C. Establish a purge gas flow of 60 to 100 ml/min overnight (sixteen hours).