Columns and Fittings

Liners/adapters and inserts, general

Table 1-3. Hardware and Recommended Fittings for Capillary Column Installation

Capillary Columns

HP Series 530 ¿

320 ¿m ID

200 ¿m ID

Metal/

 

 

 

Glass

 

 

 

 

Recommended

Capillary column

Column Fittings

nut and 1.0-mm

 

graphite ferrule, or

 

silicone O-ring(s)

Packed Column

19244-805401

Inlet Liners

(requires glass

 

insert)

Split/Splitless &

18740-60840

Split-Only

with graphite or

Capillary Inlet

silicone O-ring

Split Sampling

 

Capillary column

Same as 320 ¿m

Same as

nut and 0.5-mm

 

HP Series

graphite or

 

530 ¿

silicone O-ring(s)

 

 

Not

Not

Not

Recommended

Recommended

Recom-

 

 

mended

Same

Same

Same

Split/Splitless

18740-80220

Same

Same

Same

Capillary Inlet

with graphite or

 

 

 

Insert: Splitless

silicone O-ring

 

 

 

Sampling

 

 

 

 

Programmable

19245-205803

19245-20520

19245-20510

19245-

On-Column

 

 

 

20550

Capillary Inlet

 

 

 

 

Insert

 

 

 

 

FID/NPD2

19244-805501

Same

Same

Same

Liners/Adapters

 

 

 

 

TCD

18740-20950

19232-805501

Same

Same

Liners/Adapters

and

 

 

 

 

 

18740-209603

 

 

 

ECD

19244-805501,3

19233-80530

Same

Same

Liners/Adapters

 

 

 

 

 

 

 

 

 

 

1Use 1/4-inch swage-type nut (if a nut is not supplied as part of the adapter) and graphite or Vespel ferrule to install liner/adapter.

20.11-inch jet must be used; see Chapter 8 for information regarding jet exchange (if necessary).

3Use only if detector is not configured with capillary makeup gas adapter. If makeup adapter is provided, it is used instead (usually with makeup gas turned off).

19