Test Sample Chromatograms
Test sample chromatograms
Figure
HP 5890 Test Sample Operating Conditions
Detector Type | FIDw/MUG | ||||||
| Temp 250 DEGREES C | ||||||
Inlet Type SPLIT ONLY OR | |||||||
SPLIT/SPLITLESS Temp | 200 DEG C | ||||||
Operating Mode SPLIT(PURGE ON) | |||||||
Purge Time On | 0 |
|
| min | |||
Purge Time Off | 0 |
|
| min | |||
Oven Temp Programmed (1 ramp) | |||||||
Init Temp |
| 100 | DEGREES C | ||||
Init Time |
| 0 |
| min | |||
Ramp |
|
|
|
|
|
|
|
Rate |
|
| 20 |
|
|
| |
Fin Temp |
|
| 180 |
|
|
| |
Fin Time |
| 1 |
|
|
| ||
Range | 7 |
|
|
|
|
|
|
COLUMN: |
|
|
|
|
|
|
|
Part No. |
| 19095Z•121 |
| ||||
Dimensions |
|
| 530¿ ; 10 M | ||||
Sta Phase | Methyl Silicone |
START
2.21 2.35
2.81
FLOW RATES |
|
|
|
Carrier (He) | 15 | +/• 1 | ml/min |
Hydrogen | 30 | +/• 1 | ml/min |
Air | 400 | +/• 20 | ml/min |
Makeup (N2) | 20 | +/• 1 | ml/min |
Split Vent | 200 | +/• 20 | ml/min |
Septum Purge | 5 +/• 1 | ml/min |
SAMPLE:
Type Capillary Sample Inj Volume 1 ¿l Part No. 18740•60900 Composition: Soln of 0.1%(w/w)
.40 C9
2.64
3.67 C16
ST
230