Chapter 4---Optical
CAUTION! Fan disconnection is acceptable for a short period of time only (preferably <20
WARNING!!! Dangerous levels of ultraviolet and infrared radiation, dangerous glare, very high temperatures (180°C to 300°C) and high internal gas pressure are present at the Xenon Arc Lamp. Protect eyes from ultra violet light and infrared light by using X5 (375 to 700 nanometers), ANSI approved, shade goggles when actually working on the projector near the arc lamp source.
6.Access the "Shutters on Hide" box again from the
Green. This mutes video from the Green CRT but leaves the Green shutter open to allow Arc Lamp light from the Green ILA® to display on the screen.
7.Select ILA® Bias from the
8.Use the
9.Adjust the Arc Lamp alignment screws (see Figure
NOTE: It is easier to perform the following procedure with one person standing in front of the screen holding a light meter and another person making the adjustments to the Arc Lamp.
10.Use the light meter to locate the brightest area or “hot spot”, (see Figure
11.Adjust screw 1, 2, or 3, (see Figure
12.Check for maximum brightness and readjust the
13.Tighten spider lock down bolt.
Model 250 Service Manual |