Operation types
sensor reference point | high |
|
|
|
(flange face) |
|
|
| |
| calibration |
|
|
|
| point | distance | ||
|
|
|
| space |
|
|
|
| |
|
|
|
|
|
level
low calibration point
1.4.Units
Select the units for the Quick Start variables (high and low calibration point, and level, distance, or space).
Options | mm, cm, m, in, ft |
|
|
1.5.High Calibration Point
Distance from Sensor Reference to High Calibration Point: usually process full level. (See 1.3. Operation for an illustration.)
Values | Range | 0.0000 to 100.00 m |
|
|
|
1.6.Low Calibration Point
Distance from Sensor Reference to Low Calibration Point: usually process empty level. (See 1.3. Operation for an illustration.)
Values | Range: 0.0000 to 100.00 m |
|
|
1.7.Rate
Sets the reaction speed of the device to measurement changes in the target range.
| SLOW | 0.1 m/minute |
Options |
|
|
MED | 1.0 m/minute | |
|
|
|
| FAST | 10.0m/minute |
|
|
|
Use a setting just faster than the maximum filling or emptying rate (whichever is greater).Slower settings provide higher accuracy: faster settings allow for more level fluctuation
1.8.Apply Changes
In order to save the Quick Start settings it is necessary to enable Apply Changes.
Options | YES, NO |
|
|
Select YES. SITRANS LR 460 is now ready to operate and returns to RUN mode.
Quick Start
7ML19985JM01 | SITRANS LR 460 – INSTRUCTION MANUAL | Page 29 |