Siemens LR 460 instruction manual Environmental, Process, Communication

Models: LR 460

1 123
Download 123 pages 8.81 Kb
Page 15
Image 15

Enclosure

 

construction

Die-cast aluminum, painted (polyester powder-coated)

conduit

2 x M20

 

 

or 2 x ½” NPT (option)

ingress protection

Type 4X/NEMA 4X, Type 6/NEMA 6, IP671

Dust cap (optional)

 

3"

PTFE, pipe clamp connection, O.D. 95 mm (3.74")

4"

PTFE, pipe clamp connection, O.D. 120 mm (4.72")

Air Purge Connection

 

equipped with female 1/8" NPT fitting

Environmental

location

indoor/outdoor

altitude

2000 m (6562 ft) max

relative humidity

suitable for outdoor (Type / NEMA 4X, 6/ IP67)

installation category

II

pollution degree

4

permitted ambient

-40 to 65 °C (-40 to 149 °F) (non-hazardous version)

 

 

temperature3LCD: -10 to 55 °C (14 to 131 °F)

 

 

Observe the temperature classes in hazardous areas!

Process

Process temperature2

-40 to 200 °C (-40 to 392 °F)

Pressure (vessel)

0.5 bar (7.25 psi) maximum

Communication

Communication: HART

230 to 600 Ω, 230 to 500 when connecting a coupling

 

Load

 

 

 

 

module

 

Max. Line Length

multi-wire: 1500 m (4921 ft)

Specifications

Protocol

HART, Version 5.1

 

 

 

 

 

 

 

 

 

 

 

1.Use only approved, suitable sized hubs for watertight applications.

2.Process temperature varies with ambient temperature. See Process/Ambient de-rating curves in Appendix III.

7ML19985JM01

SITRANS LR 460 – INSTRUCTION MANUAL